A laboratory gamma-ray spectrometer built around a 1024-channel multichannel analyser and a cylindrical 3″ × 3″ NaI(Tl) scintillation detector with integrated photomultiplier. The complete detector and analyser assembly is enclosed in lead shielding with a minimum wall thickness of 85 mm, lowering the natural background by roughly two orders of magnitude to reach low detection limits. The instrument is fully controlled and powered over a single USB cable from a standard PC running the LabCenter evaluation software.
1024-channel multichannel gamma-ray spectrometer
3″ × 3″ NaI(Tl) detector with integrated photomultiplier
Lead shielding with movable lid (≥ 85 mm wall)
Controlled and powered over a single USB cable
LabCenter evaluation & reporting software
From the measured gamma spectra, LabCenter calculates concentrations of the natural radioactive elements K-40, U(nat) and Th(nat), as well as activities of man-made radionuclides such as Cs-137, Co-60 and Am-241; the range of evaluated elements can be extended. Typical applications include the steel and metallurgical industry (steel, dust and slag samples) and geophysical sample analysis. Correct activity results require geometric conformity between the measured sample and the supplied calibration standards.
| Parameter | Value | Note |
|---|---|---|
| NaI(Tl), 0.35 l, ø 75 mm × 75 mm (3″ × 3″) | Bialkali photomultiplier | |
| better than 9 % FWHM | at 662 keV | |
| 1024 | grouped from 65535 | |
| 1.5 s | - | |
| 500–1000 V | digitally controlled | |
| two-point (gain + offset) | 662 keV typ. at channel 220 | |
| automatic, 0.25 % accuracy | - | |
| USB | control & power supply | |
| Cs-137, 18 kBq | - | |
| min. 85 mm wall thickness | - | |
| 770 × 360 × 620 mm | - | |
| 580 kg | - | |
| DC +5 V, max 100 mA | via USB | |
| 0 to +40 °C | storage −20 to +70 °C |